Ultraviolet Fluorescence (UVF) Method Sulfur Content in Oil Analyzer ASTM D5453
Introduction:
The instrument is designed and manufactured according to the standards SH/T 0689-2000 Standard test method for determination of total sulfur in light hydrocarbons,motor fuels and oils by ultraviolet fluorescence,SH/T 0657-2007 Standard test method for trace nitrogen in liquid petroleum hydrocarbons by syringe/inlet oxidative combustion and chemiluminescence detection and ASTM D 5453 -2006 standard. It is widely used in petroleum,chemical industry,electric power,coal,food,environment proection and others.At present, GD-0689N is the advanced instrument to analyze the sulfur,nitrogen in China.
ASTM D5453:
It is an orthodox test method to ascertain the total sulfur in liquid hydrocarbons containing 1.0 to 8000 mg/kg total Sulfur, which boil in the range from approximately 25 °C to 400 °C and with viscosities at room temperature of between approximately 0.2 and 20 cSt (mm2/S). This method of testing also has applications in determining the total sulfur in liquid hydrocarbons which contain fewer than 0.35% (m/m) halogen(s).
Method Summary:
A measured quantity of the hydrocarbon sample is either directly injected into the pyrotube of the analyzer at a controlled speed into the vaporization and combustion zones of the instrument. In the inlet zone the sample is transported by an inert stream of a carrier gas, typically Argon or Helium. The hydrocarbon is oxidized in a rich pure oxygen stream in the combustion zone at approximately 1000°C where it is converted into CO2, SO2 and H2O. After the gases exit the combustion zone they are passed through a semipermeable Nafion® membrane where the water is quantitatively removed. The gases enter the Sulfur reaction chamber where they are exposed to an UV light. The SO2 molecules absorb the radiation and become excited. The relaxation takes a few nanoseconds releasing a secondary energy which is captured by a Photo Multiplier Tube, amplified processed and recorded. The signal obtained is proportional to the total sulfur content of the original sample.
Main technical features
1. The instrument can detect the both sulfur and nitrogen,also could test the solid sample or liquid sample.
2. With the desktop computer,Windows system,man-machine direct conversation platform,It can set all prameters and make the test by the keyboard operation.
3. It adopts the control technology of the computer to automatically collect,process,store and print all data.with the wide linearity range of measurement,strong capacity of resisting disturbance,easy operation and reliable test results.
4. With the imported fluorescence excitation source,membrane desiccator,optical filter,metal encapsulated photomultiplier, It has the features of the high sensitivity,fast,stability,high precision,good consistency.
5. With the advanced temperature control system,fast and steadily warming rate,long service life,the temperature control accuracy ±3ºC during the temperature range,It can meet the test requirements.
Technical Parameters:
Sample types: Solid, Gas and Liquid
Method of determination: Ultraviolet fluorescence method (S)
Sample injection quantity: Solid: 1-90mg; Liquid: 2-50μL; Gas: 1-50mL
Measuring range: 0.1~10000mg/L (High concentration should be diluted)
Temperature range: room temp~1300ºC
Temperature-controlled precision: ±1ºC
Sample detection time: about 2 minutes
Gas source requirements: High-purity argon: above 99.9%
High purity oxygen: above 99.9%
Power supply: AC220V±22V, 50Hz±0.5Hz, 1500 W
Dimension: Host: 305(W)×460(D)×440(H)mm
Temp controller: 550(W)×460(D)×440(H)mm
Weight: Host: 20kg; Temp controller: 40kg
Measuring accuracy:
Concentration values (ppm) |
Injection quantity (µL) |
RSD (%) |
0.2 |
20 |
25 |
5 |
10 |
10 |
50 |
10 |
5 |
100 |
10 |
3 |
5000 |
10 |
3 |
Related products
Parameter |
GD-380B |
GD-387 |
GD-17040 |
GD-0689 |
Standards |
GB/T 380,
ASTM D1266 |
GB/T 387,
ASTM D1551 |
GB/T 17040, GB/T11140, ASTM D4294 |
ISO 20846,ASTM D5453,ASTM D4629,ASTM D5762, ASTM D6069 |
Test sample |
Light oils, such as gasoline, kerosene |
Dark oils which sulfur content >0.1% |
Petroleum oil with sulfur content of 20ppm to 100% |
Petroleum oil with sulfur content of 5ppm to 100% |
Test method |
Lamp method |
Quartz tube method |
X-ray fluorescence (XRF)method |
Ultraviolet fluorescence method |
Dimension |
3800*320*600mm |
530*400*670mm |
468*368*136mm |
Host:305*460*440mm Temp controller:550*460*440mm |
Working temperature |
5~40ºC |
5~40ºC |
5~40ºC |
5~40ºC |
Relative humidity |
≤85%(30ºC) |
≤85%(30ºC) |
≤85%(30ºC) |
≤85%(30ºC) |
Power suply |
AC220V ±20V, 50Hz |
AC220V ±20V, 50Hz |
AC220V ±20V, 50Hz |
AC220V ±20V, 50Hz |
What service can our company provide you?
Pre-sale Service:
1. The professional sales team will reply to the inquiry in real time;
2. Customize suitable solutions according to the needs of different customers;
3. Provide product technical parameters/operation manuals and evaluation reports, etc.;
After-sale Service:
1.From the date of purchase within one year instrument, is a product quality problem free repair replacement, lifetime supply of maintenance and technical services.
2.Technical Support
(1)Our technical team supports our customers at all time with our long experience, expert advice and the unlimited spirit and always be for providing the best customer service.
(2)Training how to install the machine,you can arrange your engineer to our company to get free training.
(3)If necessary, we can also arrange our engineer to your company for training and commissioning at your cost.
3.The instrument has found anomalies or malfunctions please contact us to arrange the most convenient treatment options.
Packaging &Shipping:
Our product packaging has been proven over the long term and we usually have plywood boxes or wooden case with trays. We will arrange the appropriate packaging according to the weight and size of the goods. Of course, you can also choose the packaging form.
Packaging |
The product will be packed in wooden case, and also can be packed by others ways as your request. |
Both neutral and customized packing are available. |
Shipping |
We can ship the goods to you by international express,such as DHL, UPS, TNT and so on. |
And you can choose the appropriate way based on your timeline and budget. Besides, you can choose to use your own shipping agent. |
Contact Us:
Welcome to contact us talk about details if you are interested in Ultraviolet Fluorescence (UVF) Method Sulfur Content in Oil Analyzer ASTM D5453.If you want to know more information,Don't hesitation, welcome your inquiry!Price can be negotiated!