Customization: | Available |
---|---|
After-sales Service: | Online Service |
Warranty: | 12 Months |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
(1) It is composed of two sections: low temperature tempering section and its controlling section, and high temperature tempering section and its controlling section.
(2) The low temperature tempering section is equipped with a portable refrigeration device.
(3) The instrument will enter into constant temperature state automatically after setting the controlling temperature, it is not necessary to modify the set value. By the way, the temperature controller has the function of parameter modification. If there is any difference between temperature shown on the temperature controller and measured temperature on the thermometer, you can modify it.
(4) The instrument also has the function of alarm. When the temperature reaches to the temperature controlling point, you can turn on the time switch, and then the timer will time automatically and the pump will blow at the same time. It will stop blowing after 5 minutes. Stand it for 10 minutes, and the instrument will make an alarm after 10 minutes.
Power supply | AC(220±10%)V, 50Hz | |
Constant temperature heating power | 650W | |
Auxiliary heating power | 1000W | |
Cooling power | 500W | |
Air flow rate | (94±5)ml/min,adjustable | |
Temperature control range for high temperature bath | (Room temp.~99.9)ºC | |
Temperature control range for low temperature bath | (5~99.9)ºC | |
Temperature control accuracy | ±0.5ºC | |
Timer | 5min and 10min,accurate to second | |
Timing accuracy | 0.01%+0.05s(20ºC) | |
Ambient temperature | (-10~+40)ºC | |
Relative humidity | ≤85% | |
Maximum power consumption | 2700W | |
Dimension | Main unit | 800*500*765mm |
Cooler | 400*450*300mm |
No | Name | Unit | Quantity | Remark |
1 | Main unit | Set | 1 | The main unit is composed of two baths |
2 | Portable cooler | Set | 1 | |
3 | Graduated flask | Piece | 4 | 1000 ml |
4 | Thermometer | Piece | 1 | -20~102 ºC, scale division is 0.2 ºC |
5 | Foaming head (diffuser) | Piece | 4 | |
6 | Cork for graduated flask | Piece | 4 | |
7 | Drying tower | Piece | 2 | installed |
8 | Fasten device for the cover of drying tower | Piece | 2 | installed |
9 | Rubber tube | Meter | 2 | |
10 | Fuse 10 A (Φ5×20) | Piece | 2 | |
11 | Fuse 5 A (Φ5×20) | Piece | 2 | |
12 | Sensor | Piece | 2 | |
13 | Rubber stopper for thermometer | Piece | 2 | |
14 | Stopping ring for thermometer (1#) | Piece | 6 | |
15 | Fixing ring for graduated flask | Piece | 4 | "O" type ring (Ф56×5.3) |