1 | Heat capacity | 14000 J/K~15000 J/K |
2 | Measuring range | (10~35)°C |
3 | Resolution | 0.001 K |
4 | Measurement accuracy | ± 60J/g |
5 | Repeatability error | ≤0.2% (Grade C) |
6 | Pressure endurance of oxygen bomb | 20 MPa |
7 | Ambient temperature | 10 °C~28 °C |
8 | Relative humidity | ≤85% |
9 | Power supply | AC 220V±5%, 50 Hz |
10 | Dimension | 600mm×480mm×460mm |
No. | Problem | Cause | Solution |
1 | Indication light is off. | 1.No power supply. 2.The fuse is damaged. |
1.Connect the power supply 2.Replace the fuse. |
2 | Can not be ignited. | 1.The ignition electric pole is poor connected with oxygen bomb. 2.The ignition wire is short circuited. |
1.Make sure the electric pole and oxygen bomb well connected. 2.Install the ignition wire again. |
3 | Stirrer doesn't work | 1.The stirring motor is blocked. 2.The stirring motor wire is disconnected. |
1.Get rid of the blocks. 2.Connect the wire again. |
4 | Temperature doesn't decrease | 1.Water temperature in outer bucket is much lower than ambient temperature. 2.The sample mass is too less. |
1.Make determination again after regulating the outer bucket temperature. 2. 2.Use more sample to rise the temperature 2 K~3K. |
5 | Clock stops | 1. the clock has no power | 1.Replace battery 2.Manually input time to continue determination. |