Warranty: | 1 Year |
Name: | Lmited Oxygen Index Tester |
Standard: | ASTM D 2863,ISO 4589-2,Nes 714 |
Dimension: | 550(L)*435(W)*670(H)mm |
Packing Dimension: | 860(W)*670(L)*670(H)mm |
Customization: |
---|
Suppliers with verified business licenses
ISO4589-2 GB/T2406 ASTM D 2863 |
Parameters:
1. The oxygen concentration measurement range: 0~100%
2. The oxygen concentration in accuracy of ± 0.1%
3. N2 mass flow controller range: 1.0~15.0L/min, Accuracy: 2%
4. O2 mass flow controller range: 1.0~15.0L/min, Accuracy: 2%
5. Oxygen sensor response time: <2s
6. The gas flow rate in combustion tube is 40mm / s ± 2mm /s
7. The gas flow rate in top of the combustion tube is 90mm ± 10mm/s
8. The portable ignition flame can be adjusted
9. The measurement time is 5mins, and the accuracy is ± 0.5s
10. Pressure meter accuracy is ± 1%
11. Burning tube inner diameter: 80mm
12. Burning tube height: 500mm
13. Working pressure: 0.1Mpa
Operation steps of the critical oxygen index:
1. Put the size of the sample to be tested and put it in the sample bracket;
2. Install the sample bracket in the test room;
3. Adjust the test room's oxygen concentration to 0.5%;
4. Live the sample and record the ignition time;
5. Increase oxygen concentration, observe the combustion, and record the critical oxygen index.
Precautions for critical oxygen index:
1. The sample should be kept pure and not disturbed by humidity, dust, and oil pollution.
2. During operation, we need to comply with relevant safety regulations to prevent accidents such as fire;
3. During the test process, the oxygen concentration should be strictly controlled to avoid the impact on the test results.
4. After operation, clean the test equipment for the next time.