Electronic Product UL94 Horizontal Flame Testing Apparatus

Product Details
Customization: Available
After-sales Service: Yes
Warranty: 1 Year
Diamond Member Since 2007

Suppliers with verified business licenses

Audited Supplier

Audited by an independent third-party inspection agency

Registered Capital
100800000 RMB
Plant Area
>2000 square meters
  • Electronic Product UL94 Horizontal Flame Testing Apparatus
  • Electronic Product UL94 Horizontal Flame Testing Apparatus
  • Electronic Product UL94 Horizontal Flame Testing Apparatus
  • Electronic Product UL94 Horizontal Flame Testing Apparatus
Find Similar Products

Basic Info.

Model NO.
GD-UL94
Product Name
UL 94 Horizontal and Vertical Flammability Tester
Standard
UL 94
Size
1,510(W) × 630(D) × 1,400(H)mm
Power
220 V AC, 60 Hz, 1 a
Weight
60 Kg
Manual
Supplied
Utilities
Remote Controller, Methane Gas, 2 Stage Regulator,
Vent & Exhaust Fan
Open/Close, 1.4 M3/Min
Transport Package
Wooden Case
Specification
1, 510(w) * 630(D) * 1, 400(H)mm
Trademark
GOLD
Origin
China
HS Code
9031809090
Production Capacity
100 Sets/Month

Product Description

     UL 94 Horizontal and Vertical Combustion Tester
Products:

UL 94 measures the combustion properties (after flame, after glow) by burner (H-5025 UL tirrill burner) when test sample combusts in chamber's internal volume of 0.864m3 and is installed internal light to observe the specimen's properties and has ventilating opening with emission fan to remove the combustion gas.

UL 94-AVH has auto specimen moving system for protecting user from hazardous gas and auto burner moving system for meeting the whole test standards. This auto burner moving system has a goniometer , so it can change the angle of flame to the specimen (0˚, 20˚, 30˚, 45˚, 90˚).

Control box and remote controller  help user to handle this equipment or check after flame time and after glow time during testing and are able to adjust the angle. 

Electronic Product UL94 Horizontal Flame Testing Apparatus
Standard: 
 

1.Horizontal Burning Test : UL 94HB (ASTM D 5048 or ISO 10351).

2.Vertical Burning Test : UL 94 V-0, V-1, or V-2 (ASTM D 3801, IEC 707, or ISO 1210).

3.500w (125mm) Vertical Burning Test : 5VA or 5VB (ASTM D  4804 or ISO 9773).

4.Thin Material Vertical Burning Test : VTM-0, VTM-1, or VTM-2 (ASTM D 4804 or ISO 9773)

5.Horizontal Burning Foamed Material Test : HBF, HF-1, or HF-2 (ASTM D 4986 or ISO/DIS 9772.3)

6.KS M 3015 : Testing method for thermosetting plastics

7.KS M 3305 : Liquid unsaturated polyester resins for fiber reinforced   plastics

8.KS M 3809 : Thermal insulation material made of rigid urethane foam

9.KS M ISO 9772 : Cellular plastics-Determination of horizontal burning characteristics of small specimens subjected to a small flame


Specifications:
 

Shaft length 100 ± 10mm,
Inner diameter 9.5 ± 0.3mm.
The height of of flame 38 ± 2mm
The programmable controller PLC + touch screen for control / testing / computing / data display.
The volume of tank more than 0.5m³
Ambient temperature 10 ~ 35 ºC
Relative humidity ≤75%
Power supply AC 220V ± 10V   50Hz
Air purity ≥95% natural gas, propane

1. The tester consists  the test section and control components  for easy field installation and commissioning.

2. The large observation window is easy to  observe the  sample combustion state;

3. Advanced industrial appearance and control designing, easy to touch the specimen and burner;

4. Controlling the movement of the burner after closing the door.

5. Bunsen burner is in according with  ISO10093: 1998 .

Features

1.  Auto specimen moving system (The Switch 'MOVE1' in the control box enables random control and operation by user) facilitating automatic movement and removal of sample to the ignition source.

2.  Auto Burner Moving System (The Switch and Remote Controller switch 'MOVE2' in the control box enables random control and operation by user) enabling adjustment of different ignition time and burner angles by standards.

3.  Auto Burner Lifting system (The Switch and Remote Controller switch 'MOVE3' in the control box enables random control and operation by user) enabling adjustment of different height of the specimen and apply lifting flame to specimen.

4.  Cable Handy Remote Controller facilitating operation of Timer 1,2 and 3 in the UL 94-AVH Control Box and Auto Burner Moving System.

5.  Double-coated chamber equipped with exhaust fan to readily discharge combustion gas and internal lighting, whose internal volume is 0.864 m3.

6.  Burner Calibration Kit to measure 20mm flame and 125mm flame under ASTM D 5207 (Copper slug thermocouple and digital thermal indicator for 20mm flame and 125mm flame).

7.  Thermal Indicator and Thermocouple Connector in the control box facilitating calibration of flame calories by the Copper Slug(ASTM D 5207).

8.  Burner complying with ASTM D 5025 equipped with the precision control system including goniometer facilitating adjustment of burner angles(0˚, 20˚, 30˚, 45˚, 90˚), frame, and gas flow meter in addition to manometer.

9.  Control Box adjustable at intervals of 30 degrees facilitating time adjustment and operation of the timer by user's location.

10.  Sliding Door made of reinforced glass to observe sample and prevent user from inhaling noxious gases during test.

11.  2 access ports on the front of the chamber to protect user from inhaling noxious gases, adjust the burner and devices in the chamber, and move and fix samples.

12.  Clamp facilitating separation and attachment of test sample.

13.  Units and accessories to change flame shape and burner angle in diverse test methods by standards(For example, accessories capable of HBF, HF, and ISO 9772 tests).

14.  3 digital timers to measure simultaneously the time of after flame, after glow and after flame.

 

Test result: 

1.  After flame : Flame continued even after a firing source disappears   

2.  After flame Time: Time taken for after flame continued under specified conditions(unit: second)

3.  After glow: Glowing combustion continued even after an ignition source and flame disappear   

4.  After glow Time: Time taken for after glow continued under specified conditions (unit: second)   

5.  Damaged length (Damaged length of test sample, unit: mm)

6.  Ignited the cotton indicator by drip flaming(Ignition of cotton by a dropped flame piece) 

 