Customization: | Available |
---|---|
After-sales Service: | 2 Years |
Warranty: | 2 Years |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
1. Measure speed. The fastest result is 1 minute, usually 3 to 5 minutes. The results of the entire process, including injection, measurement, cleaning, and drying, are generally within 10 minutes.
2. The Houillon viscosity meter has a high degree of automation. Measurement, cleaning, drying, result calculation, and repeatability calculation are all automatically completed.
3. Multiple types of measurement samples. It can measure transparent and opaque samples, including gasoline, diesel, kerosene, cutting fluid, lubricating oil, new oil, in use oil, and other Newtonian liquids.
4. The Houillon viscosity meter measures data accurately. This instrument has a secondary precise quantification function, automatically cutting the sample liquid column, accurately controlling the sample volume, no longer relying on micro pipette quantification, avoiding the limitations and deviations of pipette measurement, and ensuring data accuracy.
5. Fully automatic dual solvent cleaning, fast cleaning, low cleaning cost. The cleaning process consumes no more than 20 milliliters of cleaning solution.
6. Auxiliary functions. Including automatic blocking, constant calibration, temperature calibration, internal clock timing check, automatic calculation of viscosity index, automatic conversion of kinematic viscosity to Ensign viscosity, and automatic calculation of repeatability.
7. Dual over temperature protection for software and hardware, over temperature alarm, and anti dry burn protection function.
8. The temperature of the viscometer rises and falls quickly. The fastest rate of temperature rise and fall is 5 ºC/minute.
9. The Houillon viscometer adopts a double-layer glass cylinder for more uniform temperature.
10. The viscometer adopts PT500 high-precision temperature sensor, and the temperature stability and accuracy of the constant temperature box are high, with a temperature control accuracy of up to 0.01 ºC.
11. The viscometer can be equipped with a disposable filter, greatly reducing the workload of experimental personnel in sample pretreatment.
12. The Houillon viscometer can be quickly replaced without emptying the thermostatic bath.
13. The viscometer can wirelessly transmit data through WIFI and connect to the LIMS system.
Timing resolution | 0.01 S |
Sample usage | 0.3~1 mL |
Repeatability | ≤0.5% |
According to the criteria | ASTM D7279 ,D445,D446,NB/SH/T 0956-2017,T/CEC 127-2016 |
Bath volume | 5.5L |
Temperature control accuracy | 0.01ºC |
Measurement range | 0.3~6000mm2/s The two viscosity tubes were measured over a span of 100 times. |
Temperature control range | 20~100ºC(Separate cooler required for target temperatures below 40°C) |
Environment temperature | 10~28ºC |