• UL1581 Vertical Wire Flame Tester
  • UL1581 Vertical Wire Flame Tester
  • UL1581 Vertical Wire Flame Tester
  • UL1581 Vertical Wire Flame Tester
  • UL1581 Vertical Wire Flame Tester

UL1581 Vertical Wire Flame Tester

After-sales Service: All Life Technical Supports
Warranty: 12 Months
Product Name: UL1581 Vertical Wire Flame Tester
Standard: UL1581, ASTM D 5025, ASTM D 5207
Power: AC220V, 50/60Hz, 1A
Manual Certificates: Supplied
Customization:
Diamond Member Since 2007

Suppliers with verified business licenses

Manufacturer/Factory, Trading Company

Basic Info.

Model NO.
GD-UL1581
Air Source
Methane Gas (Customer-Owned)
Size
2150mm X 1200mm X 2150mm
Transport Package
Plywood Case
Trademark
GOLD
Origin
Chongqing China(Mainland)
HS Code
9031809090
Production Capacity
500 Sets/ Year

Product Description

UL1581 Vertical Wire Flame Tester

Introduction:   

Cable horizontal and vertical burning test apparatus(VW-1)is designed according to UL1581 sect 1080.1~1080.14, and meet the criteria of VW-1(Vertical-specimen) Flame Test. It can be used to test of 1060 products wire and cable  FT1 and vertical burning test,1080 VW-1 test, 1061 test, 1090 horizontal test and 1100 horizontal specimen or FT2 test; to assess the flame spreading the ability to combustible nearby of cable or flexible wire when it caught fire. The whole test is carried out in a sealed test chamber with a volume of 4.32 M3.  

 

Standard:   

• UL1581 sect 1060~1100

• ASTM D 5025

• ASTM D 5207

 

Feature:   

1.  The test chamber is electrolytic plate electrostatic spraying, elegant appearance, and inside is roasted black.

2.  The tester is equipped with powerful exhaust fume extractor when the test is complete, open and air supply system and exhaust to the outdoors, and during the test, the system is closed.

3.  The air throttle is equipped behind the test chamber. After the completion of the test, open the ventilation system to air distribution in the cabinet, But during the testing, the air door should be closed.

4.  The glass viewing window is on the front box, allowing easy observation during the test.

5.  The test chamber is equipped with a sealed glove to operate the burner inside.

6.  Test chamber size: available space 2.1M (l) *1M (w) *2.1M (h), capacity: 4.48m3. Dimensions 2.4m (l) *1M (w) *2.3M (h).

 

Temperature measurement system

1.  A thermal sensor (copper tubing and thermocouples)-tube structure should be very high conductivity electrolytic copper. Length of 19.14 ± 0.02mm.

2.  K-thermocouple, 0.5mm diameter of stainless steel tubes as a sleeve.

3.  The digital temperature meter with match with imports K-type thermocouple, which readings greater than 1000 degrees centigrade.

4.  Equipped with an electronic stopwatch: accurate to 0.1s.

 

Apparatus and Fixture

1.  The main fixtures include a test stand and burner fixture.

2.  The burner inner-diameter is 9.5 ± 0.5mm, fire mouth diameter is 0.90 ± 0.03mm; burner pipe is 102mm. The air supply is adjustable.

3.  The height of the flame is adjustable; Standard requests are: total height of the flame is 125 ± 10MM, blue inner flame height is 40 ± 2 mm; flame test angle is adjustable from 0 to 45°.

4.  Imported gas special flowmeter and precise needle valve controlling gas flow.

5.  Imported brands of pressure gauges and control valve working gas pressure.

6.  The test stand is made of stainless steel, and include horizontal bracket, vertical support an independent type, also as FT1, FT2, VW-1, three kinds of specifications. The plant of vertical support and blowtorch is parallel to the observation window plane, easy for viewing and operation.

7.  The burner is imported from USA, it fit to ASTM D 5025-99 request.

8.  High voltage electronic ignition system, and automatic cut off of gas supply for security control system.

9.  Wedge-shaped frame: Add wedges, the burner is installed in the wedge, and can be easily moved and moved back to the defined position very accurately.

Product Image:
UL1581 Vertical Wire Flame Tester
UL1581 Vertical Wire Flame Tester

Specification:    

Power supply

AC220V,  50/60Hz, 1A

Manual certificates

Supplied

Air source

Methane gas (customer-owned)

Size

2150mm x 1200mm x 2150mm

 

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now

You Might Also Like

Diamond Member Since 2007

Suppliers with verified business licenses

Manufacturer/Factory, Trading Company
Registered Capital
100800000 RMB
Plant Area
>2000 square meters