Components | Detection limit | Components | Detection limit |
H2 | 2 | C2H4 | 0.05 |
CO | 2 | C2H6 | 0.05 |
CO2 | 5 | C2H2 | 0.05 |
CH4 | 0.05 |
Thermal Conductivity Detector (TCD) | Hydrogen flame ionization detector (FID) |
Semi-diffusive structure | Programmed auto-ignition (above 130°C auto-ignition) |
Power supply with constant current control | Automatic ignition; stabilization time: 10 minutes |
Sensitivity: ≥10000mV-ml/ mg (benzene) | Detection limit: ≤3x10-12g /s (n-hexadecane/isooctane) |
Baseline noise: ≤10uV | Baseline noise: ≤5x10-14A |
Baseline drift: ≤30uV/30min | Baseline drift: ≥1x10-13A/30min |
Linearity: ≥105 | Linearity: >107 |
Temperature control part | Temperature control range | Temperature Control Precision |
Column box | 4°C~450°C above room temperature | ±0.01°C |
Hydrogen Flame (FID) Detector | 4°C~450°C above room temperature | ±0.01°C |
Thermal Conductivity (TCD) Detector | 4°C~450°C above room temperature | ±0.01°C |
Ni Catalyst Converter | 4°C~450°C above room temperature | ±0.01°C |
Steps of programmed temperature increase | 20 steps | |
Program Rise Rate | 0.1~60°C/min | ±0.01°C |
Program Boost | 8 steps |