The Automatic Biodiesel Oxidation Stability Tester is manufactured in accordance with NB/SH/T 0873-2013 "Determination of Oxidation Stability of Biodiesel and Its Harmonic Fuels. Other applicable standards are as follows:
NB/SH/T 0825, EN14112, EN15751, EN16568, ISO6886, GB/T 21121.
1. The software automatically records the conductivity change curve and determines the induced inflection point by calculating the second derivative of the curve.
2. After the test is over, the results of the induction reaction time together with the test process curve will be stored in the database for the query.
3. In addition to measuring the induced reaction time, the software also allows the evaluation of the so-called settling time, that is the time required for the conductivity to reach a certain value.
4. The software can also re-evaluate each test curve. By setting the evaluation time range, a certain segment of the curve is selected for evaluation, so that some abnormal data segments due to cleaning, samples or other reasons can be avoided.
5. The database can store all test related data, including conductivity curve, induction reaction time, method used, test start and stop time, etc.
6. The test data can be queried by setting the query conditions.
7. The software also has the function of multi-curve comparison and curve analysis.
8. Display mode: curve / number, two modes
9. Temperature and flow calibration can be performed by test software
10. Work unit: four tubes
11. Each sample is individually controlled and can be set automatically according to requirements.
1. Flow range: 8~12L/h
2. Flow accuracy: ± 10%
3. Test range: 0~800 μS/cm
4. Resolution: 0.1μS / cm
5. Gas source: special diaphragm pump 10L/H
6. Temperature control method: electric heating rod
7. Working temperature: 50~150ºC
8. Temperature control accuracy: ± 0.1ºC
9. Working power: AC220V ± 10% 50HZ
10. Power: 1.1KW
11. Result reduction: computer automatic measurement and storage
Welcome to contact us talk about details if you are interested in En14112 En15751 En16568 ISO6886 Automatic Biodiesel Oxidation Stability Tester .
Pre-sale Service | 1.The professional sales team will reply to the inquiry in real time; | |
2.Customize suitable solutions according to the needs of different customers; | ||
3.Provide product technical parameters/operation manuals and evaluation reports, etc.; |
||
After-sale Service | 1.From the date of purchase within one year instrument, is a product quality problem free repair replacement, lifetime supply of maintenance and technical services. |
|
2.Technical Support |
(1)Our technical team supports our customers at all time with our long experience, expert advice and the unlimited spirit and always be for providing the best customer service. | |
(2)Training how to install the machine,you can arrange your engineer to our company to get free training. | ||
(3)If necessary, we can also arrange our engineer to your company for training and commissioning at your cost. | ||
3.The instrument has found anomalies or malfunctions please contact us to arrange the most convenient treatment options. |