Customization: | Available |
---|---|
Warranty: | 1 Year |
Power: | Electronic, AC 220V, 50/60Hz, 20A |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Consist of
Control cabinet, embedded industrial tablet computer,Test chamber.
Introduction
Critical radiation flux is the main parameter for evaluating the fire risk of floor coverings. It represents the burning and spreading ability of floor coverings under the condition of external thermal radiation, and is an important index in the classification system of fire grading of floor coverings.
Working Principle
Flooring Radiant Panel Test Apparatus covering materials is used to measure the critical heat radiation flux of floor covering materials.The critical heat radiation flux is the incident heat per unit area. The greater the incident heat is, the higher the heat required by the material is, the more stable the material is and the less likely it is to catch fire.
Structure |
Material: stainless steel; beautiful, corrosion resistance |
The radiation heat source: porous ceramic thermal radiation plate 300mmX450mm, tilt 30° |
|
Imported calorimeter: range 50KW/m |
|
The diameter of the burner: 0.7mm, 19/EA, 60 degrees from the bottom of each center line |
|
The standard plate: made of calcium silicate plate with thickness of 20±1mm and density of 850±100kg/m3 without coating layer; |
|
Small stainless steel ignition nozzle auxiliary sample ignition. 7, inlet flow control valve, precise control of propane flow |
|
With blower to provide air source, through the Venturi mixture mixing test gas |
|
System |
|
Optical path system |
The optical path system includes 2900 ± 100 K light source incandescent lamp, light source accepted as silicon photo-diode, |
imported photocell module to measure the degree of light attenuation. The system automatically calculates the transmittance and generates the transmittance curve. Free print curve |
|
Thermal radiation system |
Heat radiation pyrometer accuracy:±0.5ºC, about 1.4m away from the radiation plate |
Radiation pyrometer sensitivity : constant in the range of wavelength 1um~9um |
|
high-performance K-armored thermocouple diameter: 3.2mm with insulation and ungrounded thermal contact; 25mm under the roof of the box, 100mm after the inner wall of the box flue |
|
A radiation pyrometer with a detectable diameter range of 250mm and a black body temperature range of |
|
Within the error range, measure the heat flux from the cooling plate to 110mm ~910mm, with an interval of 110mm |
|
Control system |
Standard control cabinet with embedded integrated 10.2" touch screen |
With sound alarm and alarm accident display function. |
|
The control box has a USB/RS232 serial port, can be connected to the external desktop/laptop operation control, free print test report |
|
Programmable controller (PLC) control system |
ISO 9239-1 |
Flooring fire reaction test - Part 1: Combustion performance of flooring materials tested by radiant heat |
ISO 9239-2 |
Flooring fire reaction test - Part 2: Testing flame spread status under the heat flux is 25kW/m2 |
ASTM E 648 |
Standard Test Method for Critical Radiant Flux of Floor-Covering Systems Using a Radiant Heat Energy Source |
ASTM E 970 |
Standard Test Method for Critical Radiant Flux of Exposed Attic Floor Insulation Using a Radiant Heat Energy Source |
NFPA 253 |
Standard Test Method for using radiant heat energy, test plates overlay system Critical Radiant Flux |
GB/T 11785-2005 |
flooring combustion performance measurement - a radiant heat source |