No. | Name | Unit | Quantity | Remark |
1 | Main unit | Set | 1 | |
2 | Sample bottle | Piece | 8 | |
3 | Thermometer | Piece | 1 | 150~170 ºC, scale division is 0.5 ºC |
4 | Fuse 20 A (Φ6×30) | Piece | 2 | For later use |
5 | Socket (250 V 16 A three cores) | Piece | 1 |
No. | Problem | Cause | Solution |
1 |
The indication light is off | No power supply | Connect to power supply |
The fuse is damaged | Replace it | ||
2 |
The heater does not work | The fuse is damaged | Replace it |
The temperature controller is damaged | Replace it | ||
The heating wire is damaged | Replace it | ||
The solid relay is damaged | Replace it | ||
3 |
Great deviation between displayed temperature and measured temperature | Incorrect parameters | Modify them |
The sensor is in trouble | Replace it | ||
4 |
The rolling disk does not rotate | The motor is in trouble | Replace it |
The chain falls off | Connect the chain | ||
The screw of connection shaft is loosen | Tighten the screw | ||
5 |
No airflow | The air compressor is damaged | Replace it |
The flow meter is unstable | Adjust it | ||
The pipe of air circuit falls off | Check the pipe connection | ||
6 |
Insufficient airflow | The dry tower is not sealed | Tighten the connection parts of dry tower to seal it |
The nozzle is blocked | Clear the block using a piece of steel wire |