GDYJ-502 automatically completes the transformer oil withstand voltage test (when voltage is applied to the insulating oil, as the voltage increases, the current through the oil increases sharply to see when the original insulation performance will be lost), including boosting, maintaining, stirring, standing, calculating, printing and other operations. The tester meets IEC60156 /IS6792/BS5874, ASTM D1816, ASTM D877 test methods.
Features:
1. The price is cheaper than other 100kV Oil BDV Tester.
2. 7-inch color LCD digital display. Set and modify parameters by knob.
3. Test standards can be selected or set according to user needs.
4. High test accuracy and good repeatability of test results.
5. We will provide free PC application software for uploading test data and making WORD test report.
6. The test data can be printed by build-in micro printer.
Functions:
1 Breakdown test of user-defined
2 Proof test
3 Calibration interface for customers
4 Unloading data to PC and making word report automatically.
5 Result print
Specifications:
Output voltage |
AC 0V-100KV |
Rate of voltage rise |
0.5kv/s±5%, 2kv/s±5%, 3kv/s±5%,5kv/s±5% |
Max test times setting for user-defined mode |
9 |
Stir time setting for user-defined mode |
0-999s |
Waiting time setting for user-defined mode |
0-999s |
Accuracy |
3%RDG+0.3%FS |
Display |
256´128 LCD |
Maximum storage capacity |
100 groups of test data |
Power supply |
AC220V±10%, 50Hz/60Hz |
Working environment |
0-50ºC,≤80%RH |
Size |
620 mm*430mm*330mm |
Weight |
41kg |