Auto oil contamination particle counter for hydraulic and lubricating oil
Introduction:
- GDJ-T oil particle counter adopts the shading principle which specified by the International Hydraulic Standards Committee.
- Built-in four commonly used standards GJB 420B, NAS1638, ISO4406.
- It adopts the advanced high pressure syringe pump injection system, support custom volume.
- The particle counter is specifically designed for detecting the contaminated particles in oil, which is according to the standard GB/T 18854-2002 (ISO11171-1999) and other national and international standards.
- It is suitable for hydraulic oil, lubricating oil, power plant fuel oil, insulating oil and turbine oil.
- The device is fully compliant with the corresponding national standard and international standards.
- The oil particle counter is developed by the principle of light blockage method(shading method) which can provide fast, accurate, reliable and repeatable test results. And it can provide the complete pollution monitoring analysis report.
- The large-screen touch human-machine interface is easy to operate
Features:
- The Particle Counter adopts the counting principle of light blockage method(shading method) which specified by the International Hydraulic Standards Committee.
- It adopts the advanced injection pump combined with positive pressure injection system, so that the injection speed is stable and the sampling accuracy is high.
- Self-set sampling volume for easy detection.
- Built-in four commonly used standards GJB-420B, NAS1638, ISO4406 and ГOCT17216-71.
- Built-in two calibration curves, compatible with all domestic and international standards.
- Support for custom standard testing.
- Full-featured automatic operation without external computer or printer.
- Using laser diodes as a light source for superior performance and long life.
- It adopts the bottle sampling method, wide pressure range, can test samples of different viscosities.
- With data storage for easy viewing.
- Simple elegant design, touch screen for easy operation.
Application:
- Laboratory oil analysis
- Acceptance of hydraulic equipment and routine maintenance
- System cleaning verification
- Cleaning verification of parts
- Wear test of hydraulic components
Specifications:
Light source |
Semiconductor laser |
Particle size |
1um~100um |
Channel number |
8 channel, size can be set free(in the range of selected curve) |
Sampling volume |
0.2-1000ml |
Sampling speed |
5mL/min~80mL/min |
Resolution |
<10% |
Limit coincidence error |
10,000 grain/mL |
Calibration |
according to GB18854 |
Relative humidity |
5ml/min-80ml/min |
Ambient Temperature |
5ºC-40ºC |
Fluid Specifications |
Pressure: 0.8Mpa; Vacuum: 0.1Mpa |
Electrical Connection |
Communication: RS232 port ;
Power Supply : AC220V±10%; 50Hz; ≤200W |
Main Assembly |
the instrument and a pump |