| Customization: | Available |
|---|---|
| After-sales Service: | 2 Years |
| Warranty: | 2 Years |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
| 1 | Power supply | AC 220V (-5%~+10%), 50 Hz |
| 2 | Temperature resolution | 0.01ºC |
| 3 | Measurement range | Heating medium: distilled water +5ºC~+80ºC(softening point under 80ºC) |
| Heating medium: glycerin +32ºC~+160ºC(softening point over 80ºC) | ||
| 4 | Heating rate | Automatically adjusted to (5 ± 0.5)ºC / min three minutes after the start of the test |
| 5 | Heating power | 600W |
| 6 | Medium stirring | Magnetic stirring, the stirring speed is (0 ~ 300) rpm |
| 7 | Effective volume of beaker | 800ml |
| 8 | Data display | Touch LCD display |
| 9 | Communication interface | RS-232C |
| 10 | Ambient temperature | 15ºC ~ 35ºC |
| 11 | Relative humidity | ≤85% |
| 12 | Total power consumption | ≤700 W |
| 13 | Dimensions | 400mm×300mm×250mm |
Warning: If there is any trouble appears, please cut off the power supply immediately. Then ask a professional to check and repair it. You can use it again only after the problem is solved to avoid any accidents.
| Common Faults | Causes | Solutions |
| The Instrument dos not response after it is powered up. | Check the recoverable overload protector when the power supply is normal. | If the overload protector has been ejected, press it again. |
| False triggering of laser detection results. | 1. There are foreign objects keep out the laser line. 2. Interference from direct sunlight. |
1. Check whether the laser transmitting and receiving pipes on both sides of the bath are covered by foreign objects. 2. Move the equipment to a place where there is poor light and place it where the light cannot interfere. |
Welcome to contact us talk about details if you are interested in Asphalt Equipment Ring and Ball Apparatus for Measuring Softening Points of Bitumen