Model NO.
GD-R2223
Weight
0-100Kg
Power Source
AC220V
Reference Standard
En14112, En15751, En16568, ISO6886
Flow Range
8~12L/H
Flow Accuracy
+-10%
Test Range
0~800us/Cm
Resolution
0.1us / Cm
Gas Source
Special Diaphragm Pump 10L/H
Temperature Control Method
Electric Heating Rod
Working Temperature
50~150c
Warranty
1 Year
Transport Package
Wooden Case or Plywood
Origin
Made in China
HS Code
9031809090
Production Capacity
100sets/Month

Product Description

Auto Biodiesel Oxidation Stability Tester for testing Oxidation stability of biodiesel fuels and blends



Introduction:

The Auto Biodiesel Oxidation Stability Tester is manufactured in accordance with NB/SH/T 0873-2013 "Determination of Oxidation Stability of Biodiesel and Its Harmonic Fuels. Other applicable standards are as follows:

NB/SH/T 0825, EN14112, EN15751, EN16568, ISO6886, GB/T 21121.

The measurement principle of the instrument is that the sample is placed in a constant temperature aluminum heating block, and clean air is passed into the biodiesel sample, and the air flows through the sample into the measuring cell containing the absorption liquid and bubbling. The end of the induction period is accompanied by the production of a secondary reaction product, a volatile organic acid (mainly formic acid), which causes the conductivity in the measuring cell to accelerate. The conductivity of the absorbing liquid is recorded by the computer software, and the induction time is automatically obtained by automatically calculating the conductivity to the second-order differential maximum point of the time-varying curve.



Features:

1. The software automatically records the conductivity change curve and determines the induced inflection point by calculating the second derivative of the curve.
2. After the test is over, the results of the induction reaction time together with the test process curve will be stored in the database for the query.
3. In addition to measuring the induced reaction time, the software also allows the evaluation of the so-called settling time, that is the time required for the conductivity to reach a certain value.
4. The software can also re-evaluate each test curve. By setting the evaluation time range, a certain segment of the curve is selected for evaluation, so that some abnormal data segments due to cleaning, samples or other reasons can be avoided.
5. The database can store all test related data, including conductivity curve, induction reaction time, method used, test start and stop time, etc.
6. The test data can be queried by setting the query conditions.
7. The software also has the function of multi-curve comparison and curve analysis.
8. Display mode: curve / number, two modes
9. Temperature and flow calibration can be performed by test software
10. Work unit: four tubes
11. Each sample is individually controlled and can be set automatically according to requirements.



Parameters:

1. Flow range: 8~12L/h
2. Flow accuracy: ± 10%
3. Test range: 0~800 μS/cm
4. Resolution: 0.1μS / cm
5. Gas source: special diaphragm pump 10L/H
6. Temperature control method: electric heating rod
7. Working temperature: 50~150ºC
8. Temperature control accuracy: ± 0.1ºC
9. Working power: AC220V ± 10% 50HZ
10. Power: 1.1KW
11. Result reduction: computer automatic measurement and storage