(1) Temperature range: 300-1000 ºC, control accuracy is not greater than ± 1 ºC (duration greater than 2 min), the dynamic is not greater than ± 2.5 ºC.
(2) Ring furnace heating system showing display time-temperature curve and the actual measured temperature control charts.
(3) Temperature recorders and sheathed thermocouples.
(4) Furnace tube diameter: Ф47 + 1-1 mm, length Ф100 + 10-5 mm.
(5) quartz tube.
(6) Furnace displacement system: motor drive, speed: 10 ± 0.1mm / min.
(7) Airflow: 0-20L / min adjustable.
(8) Material smoke production rate could be automatically conversion by the system during testing.
(9) The angular velocity chart could be displayed when the rotation chart squirrel cage.
(10) The ring furnace running time is arbitrarily set.
(11) The smoke toxicity test device has a manual/automatic fast forward, and return function for the ring furnace.
(12) Each room can be switched to exhaust smoke.
(13) Flue gas collection and rationing system realize quickly mix the gas.
Control System:
Computer control system | the test is more intelligent; curve computer data could be stored and printed freely |
Temperature control thermocouple | 1mm K-type thermocouples |
The annular heating furnace temperature calibration | time, temperature curve, and the actual temperature |
The motor control system | automatically controls furnace movement; movement speed is adjustable; and also can be adjusted manually; |
Temperature range | 300-1000 ºC |
Power | 1.5KW; Heating power: 800-1000W |