I.Introduction
- GDGS Automatic 10KV Capacitance&Dissipation Factor Test Set is used to measure dissipation factor (PF) in heavy interference site such as power plants or power substations.
- GDGS Automatic Transformer Capacitance Tester, Dissipation Factor Tester can also be used in laboratory for high accurate test. The set is whole in one designed: including precision digital bridge, power unit, reference capacitor (CN), step-up transformer and other electronic circuits.
- GDGS Automatic Transformer Capacitance Tester, Dissipation Factor Tester is also using frequency shift technology, FFT (Fast Fourier Transform algorithm) and digital filtering. When started, it's running automatically and then outputs the results on LCD display and prints the results automatically.
II. Specifications:
Accuracy: Cx±(reading×1%+1pF), PF±(reading×1%+0.00040)
Interference Ratio: Interference current is no more than 2 times of test current
Capacitance Range:Internal HV 3pF~60000pF / 10kV,60pF~1uF/0.5kV, External HV 3pF~0.3uF/10kV, Resolution 0.0001pF, 4 digital
PF Range: No limit, Resolution 0.001%, C/L/R specimen test
Input current range:10μA~1A
Internal HV: 0.5~10kV / 200mA (max)
HV Adjust: Rise or fall smoothly
HV Accuracy: ±(reading×1%+10V) Resolving 1V
Frequency select: 60Hz single point, locked 55Hz/65Hz double point, locked
Test Time: 40S typical
Power Supply: 180V~270V 50Hz/60Hz auto
Operating Temperature: 0ºC~60ºC
Storage Temperature: -20ºC~60ºC
Humidity: <90%