DIN 53436 Smoke Toxicity Test Apparatus (Mouse Tester)

Product Details
Customization: Available
Standards: DIN 53436,JIS1321
Temperature Range: 300-1000 C
Diamond Member Since 2007

Suppliers with verified business licenses

Audited Supplier

Audited by an independent third-party inspection agency

Registered Capital
100800000 RMB
Plant Area
>2000 square meters
Find Similar Products

Basic Info.

Model NO.
GD-53436
Warranty
1 Year
Oversea Training
Available
Transport Package
Plywood Case
Trademark
Gold
Origin
China
HS Code
9031809090
Production Capacity
50 Sets/Month

Product Description

    DIN 53436 Smoke Toxicity Test Apparatus (Mouse Tester)
Standard:

DIN53436

JIS1321

GB/T 20285

Introduction:

The smoke toxicity test device is designed accordance to DIN53436 and it realizes smoke of burning material continuous and stable production and gas extraction, flue gas quantitative description of smoke concentration and accurate calculation of the rate of smoke production, thus it could be used for different circumstances of different materials (including different rates and different concentrations of smoke smoke production) and animal exposure evaluation at different times.

Technical characteristics:

1, Temperature range: 300-1000 ºC, control accuracy is not greater than ± 1 ºC (duration greater than 2 min), the dynamic is not greater than ± 2.5 ºC.

2, Ring furnace heating system showing display time - temperature curve and the actual measured temperature control charts.

3, Temperature recorders and sheathed thermocouple

4, Furnace tube diameter: Ф47 + 1-1 mm,  length Ф100 + 10-5 mm

5, quartz tube

6, Furnace displacement system: motor drive, speed: 10 ± 0.1mm / min

7, Air flow: 0-20L / min adjustable

8, Material smoke production rate could be automatic conversion by system during testing.

9, The angular velocity chart could be display when the rotation chart squirrel cage

10, The ring furnace running time arbitrarily set

11, The smoke toxicity test device has a manual/automatic fast forward, and return function for ring furnace

12, Each room can be switched to exhaust smoke

13, Flue gas collection and rationing system realize quickly mix the gas 

Control System:

1, Computer control system: test is more intelligent; curve computer data could be storage and print freely;

2, Temperature control thermocouple: 1mm K-type thermocouples

3, The annular heating furnace temperature calibration: time, temperature curve and the actual temperature;

4, The motor control system, automatic control furnace movement; movement speed is adjustable; and alsocan be adjusted manually;

5, The system automatically calculates the concentration of smoke production, smoke production rate; squirrel cage speed real-time display;

6, Temperature range: 300-1000 ºC;

7, Power: 1.5KW; Heating power: 800-1000W



 

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now
Contact Supplier