Customization: | Available |
---|---|
After-sales Service: | 2 Years |
Warranty: | 2 Years |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
1 |
Standard |
ASTM D445, D446, ISO3104, GB/T265, GB/T30515 |
2 |
Measurement Rang |
1. 0.3~10000mm2/s, the measuring span of the two viscosity tube is 10000 times. |
3 | Viscometer Bath Volume |
2x3L |
4 |
Sample Quantity |
2 x 16-bit |
5 |
Temperature Controlling Range |
20~100°C (Built-in Partier cooling) |
6 |
Temperature control accuracy |
0.005ºC |
7 |
Timing accuracy |
0.01S |
8 |
Repeatability |
≤0.4% |
9 |
Machine power |
≤1600 W |
10 | Host size |
A.400×700×800 (mm) B.400x700x800 (mm) |
1. Equipped with 16-bit sample tray, it can work 24 hours without interruption. And all the progress, like the intelligent sample preheating, sampling, measurement, cleaning, drying, result calculation and repeatability calculation are all automatically completed.
2. Double bath independent temperature control, double channel testing, cleaning, drying can be carried out at the same time.
3. Using photoelectric or thermal detection which can measure transparent and opaque samples, including gasoline, diesel, kerosene, cutting fluid, heat conduction oil, additives, the fresh and in used lubricating oil new oil and other Newtonian fluid.
4. It has the PTC sample intelligent preheating function which can automatically control the preheating temperature according to the preset test target temperature of the sample to be tested, and the preheating temperature range is between room temperature and 120ºC; The preheating rate is 15ºC/ min, which greatly reduces the sample constant temperature time during the test.5. With built-in partier refrigeration system, the constant temperature bath liquid is directly used as the refrigerant to cool down rapidly.
6. The application of bath liquid level adaptive balance technology (patent No. : ZL 2020 2 1531117.X), adapt to the test temperature adjustment, in the process of rising and cooling automatically increase or decrease the constant temperature medium, keep the viscosity tube immersion depth unchanged.
7. Windows operating system with all Chinese interface which is simple operation and friendly human-machine interface. It can access LIMS system and can be connected to the scan code gun, automatic input sample information.
8. The system has built-in the automatic calibration function of the viscometer constant, as well as auxiliary functions such as temperature calibration, internal clock timing calibration, and automatic conversion of kinematic viscosi-Engler viscosity and automatic calculation of viscosity index.
9. With hardware and software double over temperature protection, over temperature alarm, anti-dry burning protection, lack of cleaning liquid prompt, waste liquid cleaning prompt and other functions to ensure the safe operation of the instrument.
10. Using double glass bath design, PT1000 high precision temperature sensor and double blade stirring, the temperature of constant temperature bath is stable and accurate, and the temperature control accuracy reaches 0.005ºC.
11. Automatic intelligent selection test bulb for unknown samples through accurate suction condition algorithm, you can choose intelligent or follow.
12. With automatic intelligent double solvent cleaning function, the cleaning mode can be customized, and the sample cup can be cleaned while cleaning the viscometer, eliminating the work of cleaning the sample cup separately.
13. With a timing verification port, the instrument can be accurately verified by the standard time verification instrument.