Customization: | Available |
---|---|
After-sales Service: | Whole Life Technical Support |
Warranty: | 1 Year |
Still deciding? Get samples of $ !
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Introduction:
This Fully automatic open cup flash point tester is designed and made as per ASTM D 92 Standard Test Method for Flash and Fire Points by Cleveland Open Cup Tester. It is applicable to all petroleum products with flash points above 79ºC and below 400ºC except fuel oils.
I. Main technical features
1. It adopts an 8-inch IPS high-definition capacitive screen. Using 32-bit ARM processor and high-precision AD chip, the test data and parameters are recorded in real time and displayed by curve.
2. Automatic measurement of atmospheric pressure, automatic correction of flash point value.
3. Equipped with Bluetooth interface.The user can query test data anytime.
4. Electronic ignition, gas flame..
5. Remote automatic upgrade to obtain the latest version, remote prediction, abnormal early warning, comprehensive evaluation and maintenance of instrument operation status.
II. Main technical specifications
1. Power supply: AC(220±5%)V, 50Hz
2.Total power consumption:<=700W
3. Flash point determination: Range:Ambient to 400ºC;Accuracy:0.1ºC
4. Heating rate: (1)Initial heating:(14~17)ºC/min
(2)It is (5-6)ºC/ min after reaching the preset flash point 20ºC
5. Igniting mode:(1)Electric ignition,
(2)Diameter of igniter:0.7mm~0.8mm
6. Fire extinguishing device: (1) when the fire point appears, it can automatically extinguish the fire and return to its original position; (2) it can also manually press the key to extinguish the fire.
7. Data transmission mode:Bluetooth
8. Ambient temperature:(15~35)ºC
Relative humidity:≤85%
9.Dimension:510mm×320mm×330mm
10. Weight:20kg
GD-3536 Series ASTM D92 CLEVELAND OPEN CUP FLASH POINT TESTER:
Parameter |
GD-3536 |
GD-3536-1 |
GD-3536A |
GD-3536D |
Temperature range |
Ambient temperature~400ºC |
Ambient temperature~400ºC |
Ambient temperature~400ºC |
Ambient temperature~400ºC |
Ignition mode |
Ignition once manually |
Ignition once manually | Ignition once manually |
Automatic |
Temperature display |
Thermometer |
LCD |
Color LCD |
Color LCD |
Printing |
No |
No |
Yes |
Yes |
Communication port |
No |
No |
No |
Yes |
Other function |
Manually adjust heating rate |
Automatically control temperature and display by LCD |
Automatic |
Automatic, with RS232 communication port |
Dimension |
350*290*350mm |
340*320*450mm |
410*360*420mm |
520*380*310mm |