Customization: | Available |
---|---|
After-sales Service: | Lifetime Technical Support |
Warranty: | 1 Year |
Still deciding? Get samples of US$ 8500/Piece
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Standards | SH/T 0794 ASTM D5191 |
Pressure measuring range | 7~200kPa Resolution 0.01kPa |
Specimen usage | 3mL |
Temperature control accuracy | 37.8ºC±0.1ºC |
Gas-tight syringe | 5mL |
Single test time | 5 minutes. |
Vacuum pumps | Vacuum level 1Pa achieved |
Display | inch color LCD display unit Resolution: 1024*600 |
Dimensions and weight | W330* D220 * H310 mm 10 kg |
Power supply and power consumption | AC 220V ±5%;200w |
System Support | LIMS system, GD-Lab system |
Vapor Pressure Test Method:
The main standard methods for vapor pressure determination are the tertiary expansion method, the micro method, and the Reid method.
1. Petroleum products as well as chemical reagents can be tested for vapor pressure by the tertiary expansion method. The method refers to standards such as SH/T 0769, SN/T 2932 and ASTM D6378 (this standard applies to the determination of temperature 0 ºC ~ 100 ºC, the gas-liquid ratio of 4:1 ~ 0.02:1, the vapor pressure is not higher than 500kPa volatile liquid chemicals vapor pressure determination).
2. The vapor pressure of petroleum products can also be measured by the trace method. The method refers to ASTM D5191 and SH/T 0794 (this standard applies to the boiling point higher than 0 ºC, and in 37.8 ºC, gas-liquid ratio of 4:1 conditions under the vapor pressure of 7kPa ~ 130kPa of liquid petroleum products, and the sample size required in 1ml ~ 10ml)
3. The Reid method is also often used for the testing of vapor pressure of petroleum products. The method refers to GB/T 8017, ASTM D323 and other standards.
Saturated Vapor Pressure Selection List:
Product details: