Customization: | Available |
---|---|
After-sales Service: | Lifetime Technical Support |
Warranty: | 1 Year |
Still deciding? Get samples of US$ 5/Set
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Constant Temperature Kinematic Viscosity Bath Kinematic Viscometer ASTM D445
The basic structure and characteristics of kinematic viscometer
Product name | kinematic viscosity bath ASTM D445 |
Power supply | AC(220±10%)V, 50Hz±5% |
Heating power | 1600W |
Stirring motor | 6W, 1200RPM |
Temperature control range | Ambient to 100ºC |
Temperature control accuracy | ±0.1ºC |
Constant temperature bath | 20L,double shell structure |
Working environment | Ambient temperature: room temperature~35ºC Relative humidity: ≤85% |
Temperature sensor | RTD, Pt100 |
Maximum power consumption | 1800W |
Capillary viscometer tubes (Pinkevitch viscometer) | 7 pieces in total, inner diameter for each: 0.6mm, 0.8mm, 1.0mm, 1.2mm, 1.5mm, 2.0mm, 2.5mm |
Dimension | 530*400*670mm |
Note: This Kinematic Viscometer Bath is equipped with 7 pieces of Pinkevitch viscometer tubes. We can customize instrument equipping with Cannon-Fenske Opaque(or Routine)Viscometer tubes according to customer's requirements.
Detail Images:
Normal problem and solution
No. | Problem | Cause | Solution |
1 | The indication light of power is off | The indication light is damaged | Replace the power switch |
The fuse is damaged | Replace it | ||
No power | Check the outer power supply | ||
2 | The outer shell is with electricity | The instrument is bad grounded | Check the grounding wire |
3 | The heater does not work | The silicon controlled rectifier is damaged | Replace it |
The heater is damaged | Replace it | ||
4 | There is nothing shown on the PV window | The temperature sensor is damaged | Replace it |
The temperature sensor is bad connected | Connect the temperature sensor | ||
5 | Over-regulation value is more than 0.2 ºC | Please set the room temperature of temperature controller again | Set it again |
Warning: If there is any trouble appears, please cut off the power supply immediately. Then ask a professional to check and repair it. You can use it only after the problem is solved to avoid any accidents.
Petroleum & Oil Analyzer Petroleum Testing Instruments
Parameter | GD-265C | GD-265D-1 | GD-265G | GD-265H |
Auto./Manual | Manual | Semi-automatic | Manual | Automatic |
Temperature range | Ambient~100ºC | Ambient~100ºC | Ambient~-70ºC with compressor for refrigeration | Ambient~100 ºC |
Temp.control precision | ±0.1ºC | ±0.01ºC | ±0.1ºC | ±0.01ºC |
Temp.control method | Continuous | Continuous | Continuous | Continuous |
Capillary viscometer | Inner dia. 0.6, 0.8, 1.0, 1.2, 1.5, 2.0, 2.5mm each one ubbelohde capillary viscometer and Cannon-fenske counterflow capillary visvometer is available for option | |||
Single/double glass tank | Double | Double | single | Double |
Sample quantity | 4 | 4 | 2 | 4 |
Timing function | No | Yes | No | Yes |
Auto-calculate and print | No | No | No | Yes |
Size | 530*400*670mm | 530*400*670mm | 530*460*870mm | 530*400*670mm |
Contact Us:
If you have interested in Constant Temperature Kinematic Viscosity Bath Kinematic Viscometer ASTM D445 , please feel free to contact with me:
Chongqing Gold Mechanical&Electrical Equipment Co., Ltd
website: chinainstrument.en.made-in-china.com