ASTM D4294 Fully Automatic Low Detection Automatic Sulfur in Oil Analyzer

Product Details
Customization: Available
After-sales Service: Online and Offline Service
Warranty: 2 Years
Still deciding? Get samples of US$ 9000/Piece
Request Sample
Diamond Member Since 2007

Suppliers with verified business licenses

Audited Supplier

Audited by an independent third-party inspection agency

Registered Capital
100800000 RMB
Plant Area
>2000 square meters
  • ASTM D4294 Fully Automatic Low Detection Automatic Sulfur in Oil Analyzer
  • ASTM D4294 Fully Automatic Low Detection Automatic Sulfur in Oil Analyzer
  • ASTM D4294 Fully Automatic Low Detection Automatic Sulfur in Oil Analyzer
  • ASTM D4294 Fully Automatic Low Detection Automatic Sulfur in Oil Analyzer
  • ASTM D4294 Fully Automatic Low Detection Automatic Sulfur in Oil Analyzer
  • ASTM D4294 Fully Automatic Low Detection Automatic Sulfur in Oil Analyzer
Find Similar Products

Basic Info.

Model NO.
GD-17040
Max Diameter of Workpiece
<500mm
Mass Scope of Workpiece
<900kg
Type
Universal Testing Machine
Weight
0-100Kg
Power Source
AC220V
Total Sulfur Analyzer
Automatic Sulfur in Oil Analyzer
Application Field
Petroleum Products, Coal, etc.
Sulfur Analyzer
ASTM D4294, GB/T11140, GB/T 17040
Xrf Sulfur Analyzer Measurement Range
Sulfur Content From 17ppm~5%
Xrf Sulfur Analyzer Test Mode
Automatic
Sample Quantity
2-3ml
Price
Competitive
Packing Size
51*43*27cm, 54*49*29cm
Gross Weight
23kg
Warranty Period
1 Year
Transport Package
Packed in Veneer Case
Specification
Refer to detailed description
Trademark
GOLD
Origin
Chongqing, China
HS Code
9031809090
Production Capacity
100 Sets/ Month

Product Description

ASTM D4294 Fully Automatic Low Detection Automatic Sulfur in Oil Analyzer
 
 
Specifications:
No. Parameters Specifications
1 Measurement range: 17 ppm to 5%
2 Repeatability <0.029 ( S+0.6)
Reproducibility <0.063 ( S+0.6)
3 Sample quantity 2~3ml (equal to sample depth of 3~4 mm)
4 Measurement time 60, 120, 240, 300, and 600 s at random
5 Calibration curve numbers 9. 5 of them are linear equation in one unknown and 4 of them are binomial parabola
6 Measurement times It can set 2, 3, 5, 10, and 50 times at random
7 Working Temperature 5~40C
8 Relative humidity ≤85% (30C)
9 Power supply AC 220V±20V, 50 Hz
10 Rated power 30 W
11 Size and weight 468 mm*368 mm*136 mm; 13 kg

 
Application:
1. The Sulfur in Oil Analyzer can be used to determine mass% of total sulfur in the crude petroleum, petroleum, heavy oil, diesel oil, gasoline, and naphtha.
2. The sulfur analyzer can be used to determine total sulfur content in the products of coal chemical industry, such as primary benzene.
3. It can be used to determine total sulfur or sulfide in other liquid or solid powder samples.
 

Introduction:
1. The xrf sulfur in oil analyzer is specially designed to determine sulfur content in the petroleum products. 
2. Adopts energy-dispersive principle, and with integrated design, it is rapid and accurate. 
3. It can meet the requirements of National Standard GB/T 17040 and GB/T11140 on repeat-ability and reproduction, and it can also meet requirements of ASTM D4294-03 on these indexes. 
4. The total sulfur analyzer provides a measure to determine sulfur content during petroleum or petrochemical production process.
 
 
Characteristics:

1. Electromechanical integration; microprocessor control; large screen LCD display; Man-machine conversation; simple and beautiful;
2. The Sulfur in Oil Analyzer can determine various products and in a wide measurement range. It is rapid for analysis and it only need little standard sample.
3. Adopts fluorescence intensity ratio analysis methods, it can make correction to temperature and pressure automatically and 
it can also make correction to ratio of carbon and hydrogen (C/H).
4. The total sulfur analyzer has self-diagnostic function, so it can determine its working state and electric parameters.
5. It takes disposable sample cell with Mylar film, so it can avoid cross contamination. Sample cell is made by a multi-function pressure device, so it is rapid and convenient. It can record sample serial number, S% (m/m), and date.
6. The sample holder can be precisely located. It can move out from inside of the instrument for sample location and anti-leakage parts replacement. It is convenient and it can avoid any contamination to detection system.
7. It can save large quantity of test data. You can browse test data and calibration curve at any time.
8. It uses a thermal printer, so it is easy to replace printing paper. It is needed to speed money on printing ribbon.
9. Equipped with a RS232 serial port, it can connect with any computer or web system.
10. Its safe X-ray protection measures can keep people from injury of X-ray radiation.


Real Pictures of ASTM D4294  XRF Sulfur in Oil Analyzer for xrf sulfur analysis

ASTM D4294 Fully Automatic Low Detection Automatic Sulfur in Oil Analyzer

ASTM D4294 Fully Automatic Low Detection Automatic Sulfur in Oil Analyzer


 

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now
Contact Supplier