Still deciding? Get samples of $ !
Request Sample
|
| Customization: | Available |
|---|---|
| After-sales Service: | 2 Years |
| Warranty: | 2 Years |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
| No. | Parameters | Specifications |
| 1 | Measurement range: | 17 ppm to 5% |
| 2 | Repeatability | <0.029 ( S+0.6) |
| Reproducibility | <0.063 ( S+0.6) | |
| 3 | Sample quantity | 2~3ml (equal to sample depth of 3~4 mm) |
| 4 | Measurement time | 60, 120, 240, 300, and 600 s at random |
| 5 | Calibration curve numbers | 9. 5 of them are linear equation in one unknown and 4 of them are binomial parabola |
| 6 | Measurement times | It can set 2, 3, 5, 10, and 50 times at random |
| 7 | Working Temperature | 5~40C |
| 8 | Relative humidity | ≤85% (30C) |
| 9 | Power supply | AC 220V±20V, 50 Hz |
| 10 | Rated power | 30 W |
| 11 | Size and weight | 468 mm*368 mm*136 mm; 13 kg |
1. Electromechanical integration; microprocessor control; large screen LCD display; Man-machine conversation; simple and beautiful;