Customization: | Available |
---|---|
After-sales Service: | All Life Technical Supports |
Warranty: | 1 Year |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
ISO 1182 Non-Combustibility Test Machine for Building Material
Introduction:
Non-Combustibility Tester measures the incombustibility and combustibility of uniform building material and non-uniform building material of actual element.
The evaluations (results) are rising temperature in the furnace, basic mass loss, and ignition of the specimen.
It fully satisfies KS F 2271, KS F ISO 1182, and ISO 1182 as international standards at the same time.
And another advantage is the automatic up-down specimen movement device for accurate results as well as the user-friendly process. Thus, it doesn't matter who is test-users.
RS 232C communication and Data Acquisition Board is able to be an accurate Interface of test values, so users can test the whole function and test by PC easily and automatically.
Application:
1. Building Materials
2. Rail
3. IMO
Test standard:
ISO 1182:2020: Reaction to fire test for building material - Non-combustibility test
BS 476-4: Fire Tests on Building Material and Structures. Part 4: Non-combustibility Test for Materials
BS 476-11: Fire test on building materials and structures - Part 11: Method for assessing the heat emission from building materials.
Feature:
1. Single Zone Furnace operable at up to 900 ºC with the NIKROTHAL Ribbon Heater
2. 750±5ºC temperature auto control with TPR
3. Furnace facilitating maintenance and replacement
4. Up/down automatic moving system for correct positioning of the test sample and user's convenience
5. Mirror to observe the ignition quality of the test sample
6. RS 232C communication and DAQ Board-adopted interface
7. Auto testing by DAQ Board and PC
8. Alarm temperature-based over current shut-off system for user's safety
9. Software (Non-Comb 2005) to print out reports based on exact data results and each standard
Test Result:
* Mass Loss(g) and Rate(%)
* Ignition Time(s) and Sustained Flaming(s)
* Initial & Maximum Furnace Temperature(ºC)
* Final Furnace Temperature(ºC)
* Maximum Specimen Center Temperature(ºC)
* Final Specimen Center Temperature(ºC)
* Furnace Temperature Rise(ºC)
* Specimen Center & Surface Temperature Rise(ºC)
* Differential Test Result by Standards
(ISO 1182:1990, KS F ISO 1182: 2004, IMO FTPC Pt.1: 1998, KS F 2271: 1998)
Specifications:
Size | 400mm(w) × 400mm(D) × 1,270mm(H) |
Power | 220 V AC, 60 Hz, 10 A |
Weight | about 40kg |
Control unit | 530mm(w) × 270mm(D) × 280mm(H) |
Manual | supplied |
Utilities | Personal Computer and PC Printer |